Eberli FR, Oldroyd KG, Urban P, Krucoff MW, Morice MC, Tanguay JF, Leon MB, Brunel P, Maillard L, Lipiecki J, Cook S, Berland J, Hovasse T, Carrié D, Schütte D, Sadozai Slama S, Garot P
DOI: 10.1136/openhrt-2024-002679
Abstract
Background
For high bleeding-risk patients (HBR) undergoing percutaneous coronary intervention (PCI), the LEADERS FREE (LF) and LEADERS FREE II (LF II) trials established the safety and efficacy of a stainless steel polymer-free biolimus-coated stent (SS-BCS) with 30 days of dual antiplatelet treatment (DAPT). The LEADERS FREE III (LF III) trial investigated clinical outcomes after PCI with the next-generation cobalt-chromium thin-strut polymer-free biolimus-coated stent (CoCr-BCS) in HBR patients.
Objectives
To report the final 3-year results of the LF III trial and compare them to LF II.
Methods
LF III was a prospective, multicentre, open-label single-arm study to evaluate the safety and efficacy of the CoCr-BCS stent. The primary safety endpoint was the composite of cardiac death (CD), myocardial infarction(MI) or definite/probable stent thrombosis (ST). The primary efficacy endpoint was clinically driven target lesion revascularisation (cd-TLR). We performed a propensity-matched comparison to the 3-year outcomes of LF II.
Results
After 3 years, CD/MI/ST had occurred in 57 patients (15%, 95% CI 11.8% to 19%) and cd-TLR in 23 (6.2%, 95% CI 4.1% to 9.2%) patients. In a propensity-matched comparison of patients treated with the CoCr-BCS versus the SS-BCS, there were similar rates of CD (6.6% vs 7.8%, p=0.50), MI (7.1% vs 8.3%, p=0.47) and definite/probable ST (1.1% vs 2%, HR 0.56, 95% CI 0.16 to 1.93, p=0.35). The rates of cd-TLR were 5.3% with CoCr-BCS versus 9.8% with SS-BCS (HR 0.54, 95% CI 0.31 to 0.96, p=0.03).
Conclusions
LF III confirms the long-term safety and efficacy of the CoCr-BCS in HBR patients treated with 1 month of DAPT.